Product

AvCount Lite Particle Counter for Skydrol - SA1805-2

Methods
ASTM D975
ASTM D6786
ASTM D7619
IP 565
IP 630
DEF STAN 91-091
DEF STAN 91-86
IEC 60970
ISO 4406
JIS B 9932:2012
JIS B 9933:2021
JIS B 9934:2012

The AvCount Light for Skydrol has been specifically designed to measure the particulate of Skydrol hydraulic fluid.

Skydrol is an advanced fire resistant aviation hydraulic fluid which inhibits corrosion and prevents erosion damage to servo valves. It has been approved by most airframe manufacturers including Airbus, Boeing and BAE Systems and has been used in their products for over 40 years.

Due to the highly corrosive nature of Skydrol material this special version of the AvCount is fitted with Isolast seals and a protective inner case.

With a simple test procedure and sub 4 minute test time, operator skill or expertise is kept to a minimum.

AvCount Lite can hold up to 3 test methods with custom methods created on a PC before being uploaded.

Test progress and results for up to 6 size bands are displayed as cumulative particles/ml and ISO 4406 cleanliness codes.

Supplied with: sample bottles, sample probe, waste pipe, universal voltage power supply, USB cable, calibration certificate and a USB memory stick containing handbooks and Partikel software for basic test method programming and data download. Learn more about Partikel Software >

Why measure particle counting in diesel fuel?

IP 630, Seta AvCount Particle Counters now a mandated test in the new EN 590 Diesel fuel specification >

AvCount Lite Particle Counter for Skydrol - SA1805-2

Features

  • ISO 11171 calibration
  • Cumulative Particles/ml
  • ISO 4406 Cleanliness Codes
  • Portable, compact instrument
  • Online/offline (bottle sample)
  • Stand-alone or PC controlled
  • Programmable via PC
  • Integrated printer
  • Battery power optional

Information

Details

Sample viscosity:
The AvCount Lite can analyse samples with kinematic viscosities of up to 64 mm2/s. When used in conjunction with the optional Sample Delivery System, the viscosity range is extended up to 200 mm/s.

Precision and accuracy:
AvCount Lite is fitted with a precision dual piston pump. When measuring samples from a bottle, the pump draws an aliqot of sample through the measurement cell at a consistent flow rate. When connected to a pressurised line, the pump provides accurate metering and control.

Protrend Software:
Optional Protrend Software SA1810-0 can be purchased to control an AvCount Lite from a personal computer or laptop. Local custom test methods can be created, tests run and data downloaded. Protrend supports multiple AvCount Lite units, synchronising the sampling and measurement to allow trend analysis.

Calibration and verification:
AvCount Lite utilises 16 point calibration. Calibration can be carried out either using NIST traceable standard solutions or by interfacing with a primary calibrated ‘Master’ AvCount Lite. Calibration can be carried out at a user site or at our authorised regional laboratories. Seta authorised training is required to perform a calibration. Verification can be performed using AvCount Verification Material.

Details

DEF STAN 91-091 Aviaton Fuel Specification
ASTM D975 Standard Specification for Diesel Fuel Oils , biodiesel, biodiesel blend, diesel, fuel oil, petroleum and petroleum products
IP 565 Determination of the level of cleanliness of aviation turbine fuels – portable automatic particle counter method
ASTM D7619 Method for counting and sizing particles in light and middle distillate fuels, and liquid bio fuels, by automatic particle counter
DEF STAN 91-86 Turbine Fuel, Aviation Kerosine Type: High Flash Type
ISO 4406 Hydraulic fluid power - Fluids - Method for coding the level of contamination by solid particles
ASTM D6786 Standard test method for particle count in mineral insulating oil using automatic optical particle counters
IEC 60970 Insulating Liquids, Methods for Counting and Sizing Particles
JIS B 9932:2012 Calibration method of automatic particle counter for hydraulic pressure-liquid
JIS B 9933:2021 Hydraulic-solid particle contamination degree code display
JIS B 9934:2012 By automatic particle counting method using hydraulic pressure-light shielding principle Cleanliness measurement method
IP 630 Determination of the concentration of dispersed particles in diesel fuel – Automatic Particle Counter (APC) Light Obscuration Method

Details

Particle Size RangeISO 11171: 4 µm(c) to >100 µm(c) (calibration for larger sizes available on request)
Number of Measuring Channels6 size channels displayed on instrument, up to 16 programmable via PC 4 µm(c), 6 µm(c), 10 µm(c), 14 µm(c), 21 µm(c), 25 µm(c), 30 µm(c), 38 µm(c), 70 µm(c)
Counts Per Measurement (Max)60,000 per ml
Sample Viscosity (Max)64 mm2/s (using internal pump), 200 mm2/s (pressure fed @ 3 BarG) (SA1950-0 Sample Delivery System is available as an accessory)
Sample Temperature RangeAmbient 0 to 70 °C
Results900 result memory Date/time indexed (last 20 tests can be viewed on screen) Download to PC/laptop port or ProTrend software and print via internal printer
Sample Flow Rate30 ml/min ±5 ml/min
Total Sample Volume Used80 ml for ASTM D7619 & IP 565, from 20 ml for other methods (includes flush cycles)
Voltage12 Vdc 3A with printer on, 1A with printer off Charger/mains adaptor 100 to 240 Vac, 50/60Hz input, 12 Vdc 2.5A output Optional 12 Vdc battery and charger (SA1816-0
Size (HxWxD)25 cm x 14.5 cm x 35 cm
Weight7.5 kg
CCCN CodeTariff 90278990

Methods

ASTM D975
ASTM D6786
ASTM D7619
IP 565
IP 630
DEF STAN 91-091
DEF STAN 91-86
IEC 60970
ISO 4406
JIS B 9932:2012
JIS B 9933:2021
JIS B 9934:2012

Download product profile

To view our full range of products, download our latest Product Profile here.

Product profile image

Want to know more?

View our full range of products, or simply get in contact with us for more information.

Features equipment