Product

SetaCool-20 - 94165-0

Methods
ASTM D97
ASTM D2500
ASTM D5853
ASTM D6371
IP 15
IP 219
IP 309
IP 441
BS 2000-15 & ISO 3016
BS EN 116 & BS 2000-309
BS EN 23015 & BS 2000-219 & ISO 3015

The SetaCool-20 is designed to provide fast, accurate and precise determination of Cloud Point, Pour Point and Cold Filter Plugging Point of petroleum products, diesel and cooling fuels to -78 °C.

Five independent units, each with four test jackets allows for easier maintenance and servicing. The instrument features a built in cooling system and dry block bath which means no external cooling medium is required, whilst a built in heated top plate prevents ice formation.

A state of the art cooling system eliminates the use of compressors and HFC (greenhouse gas) or flammable refrigerants, providing a safe and environmentally sustainable solution.

SetaCool-20 - 94165-0

Features

  • Ambient to -78 °C
  • Temperature stability +/- 0.2 °C or better
  • Built in cooling
  • 5 independent units x 4 test jackets
  • Low power consumption

Information

Details

The SetaCool-20 is operated via a simple to use digital controller and consists of a dry block bath with five units x four test jackets. The temperature is set and sample is cooled as specified by the test method. A thermometer well is used to verify the bath temperature is correct. Test results are observed as specified by the test method.

For safety, the SetaCool-20 consists of a heated top plate to prevent condensation and ice forming when the bath is cooling.

Details

ASTM D5853 Standard test method for pour point of crude oils
IP 309 Diesel and domestic heating fuels - Determination of cold filter plugging point
ASTM D6371 Standard test method for cold filter plugging point of diesel and heating fuels
IP 219 Determination of cloud point
ASTM D97 Standard test method for pour point of petroleum products.
ASTM D2500 Standard test method for cloud point of petroleum products and liquid fuels
IP 15 Petroleum products, determination of pour point.
IP 441 Determination of the pour point of crude petroleum.
BS 2000-15 & ISO 3016 Methods of test for petroleum and its products. Determination of pour point
BS EN 23015 & BS 2000-219 & ISO 3015 Petroleum products. Determination of cloud point.
BS EN 116 & BS 2000-309 Diesel and domestic heating fuels. Determination of cold filter plugging point. Stepwise cooling bath method

Details

Number of test jacket4 places x 5 units
Cooling temperature rangeAdjustable between Ambient to -78 with PID control
Temperature stability+/- 0.2 or bettwe
Temperature display resolution1 decimal place
Power100 V to 240 Vac 50/60 Hz
Size (HxWxD)Individual unit: 45 x 32 x 24.9 cm
WeightIndividual unit: 19.4 kg
CCCN CodeTariff 90262020

Download product profile

To view our full range of products, download our latest Product Profile here.

Product profile image

Want to know more?

View our full range of products, or simply get in contact with us for more information.