+44 (0)1932 564391

sales@stanhope-seta.co.uk
Find your local representative

Setaflash Series 8 Flash Point Tester HighTemp - Gas Ignitor

Part Number 82050-0

The Setaflash Series 8 is an automated closed cup flash point tester which operates with gas ignition across the temperature range +5 to 300 C.

There are two modes of operation:
- Flash / No Flash
- Ramp

The ramp mode increases the sample temperature at a rate of 2 C/min and automatically dips the ignitor every 1 C up to 100 C (every 2 C above 100 C), until either a flash is detected or the end of test temperature is reached.

ADDITIONAL INFORMATION

The automatic flash point detection system is tuned to ignore halo and other false flash effects, enhancing precision and accuracy. Barometric pressure is measured by an internal barometer, and the correction is automatically applied to the test result.


Peltier cell technology is used for heating and cooling, ensuring precise temperature control and rapid cooldown after a test is completed. The bright colour LCD display shows test parameters and the flash point test result. The text and colour of the top line indicates instrument status. All functions are accessed via menus, navigated by the keypad. Up to 64 test results can be stored in memory, each tagged with a sample reference and operator name. Results can be downloaded to a PC or printer via the RS232C port. The firmware is upgradeable for future developments.


SUPPLIED WITH: Ignitor, 2 ml syringe, mains lead and instruction manual.

SPECIFICATIONS

Temperature Range Ambient +5 to 300 C
Sample Size 2 or 4 ml according to method
Ignitor Gas
Test duration, Rapid Equilibrium Mode 1 minute below 100 C, 2 minutes above 100 C, FAME 1 minute or user defined 1 to 99 minutes
Test duration, Ramp Mode Typically 7 minutes
Heating / Cooling Method Ceramic Pad, Forced Air (post-test cooldown)
Sample Cup Material Aluminium
Voltage 100 to 250 V, 50/60 Hz
Power 300 W
Size (HxWxD) 30 x 34 x 38 cm
Weight 8 kg
CCCN Code: Tariff 90268020

DETAILS OF METHODS

ASTM D3278Standard Test Methods for Flash Point of Liquids by Small Scale Closed-Cup Apparatus - paints, enamels, lacquers, varnishes, and related products
ISO 3680 (Obs)Methods of test for paints. Tests on liquid paints (excluding chemical tests). Test for flash/no flash (rapid equilibrium method)
ASTM E502Standard Test Method for Selection and Use of ASTM Standards for the Determination of Flash Point of Chemicals by Closed Cup Methods
ISO 3679Determination of flash point. Rapid equilibrium closed cup method
ASTM D7236Flash Point by Small Scale Closed Cup Ramp Method
IP 534Flash Point by Small Scale Closed Cup Ramp Method
ISO 9038Determination of sustained combustibility of liquids
CLP Regulations EC No 1272/2008European Regulation (EC) No 1272/2008 on classification, labelling and packaging of substances and mixtures
ASTM D8174Standard Test Method for Finite Flash Point Determination of Liquid Wastes by Small-Scale Closed Cup Tester
IP 523Determination of flash point. Rapid equilibrium closed cup method
IP 602Determination of sustained combustibility of liquids
EPA 1020 A & BSetaflash Closed-Cup Method for determining ignitability
ASTM D3828Standard Test Methods for Flash Point by Small Scale Closed Tester
High Temp Gas Lid

FEATURES

Ambient +5 to 300 C
Small sample size, 2 or 4 ml
Flash/No Flash & Ramp modes
Gas Ignitor
Automatic dipping and flash detection
Forced Air Cooling
64 Test memory & RS232 interface
C or F temperature display

METHODS

ASTM D3278, ASTM D3828, ASTM D7236, ASTM D8174, ASTM E502, CLP Regulations EC No 1272/2008, EPA 1020 A & B, IP 523, IP 534, IP 602, ISO 3679, ISO 3680 (Obs), ISO 9038

INDUSTRIES

Petroleum - Fuels & Oils

Crude Oil

Aviation

Energy & Power

Chemicals & Solvents

Cosmetics & Fragrances

Marine

Automotive

Transportation, Storage & Packaging

Paint, Varnish & Waxes

Pharmaceuticals

Printing Inks

Waste & Environmental

Biodiesel

Lubricants

Copyright Stanhope–Seta Limited 2014